eSeasongear SALT-3050 Re-calibration Instruction

eSeasongear SALT-3050 Re-calibration Instruction
eSeasongear SALT-3050 Salt Meter Re-calibration Instruction













1. Clean the probes with isopropyl alcohol and a cotton swab.

And clean the probe with diluted Muriatic Acid if available.

2. Press [POWER/ENTER] to turn the meter on. The reading will always be 0 ppm in the air.

3. Make sure the meter is at 0.5 ppm mode. Press [HOLD/MODE] until mode switches to 0.5 ppm if needed.

- New 2023 SALT-3050 is preset to the 0.5 mode already, so there is no need to set mode any more.

4. Insert the meter into known value calibration solution about 1-2 inches (at room temperature 25C / 77F).

5. Lightly swirl the meter to ensure the removal of air bubbles or electric charges. Wait until the reading stabilizes (10-30 seconds),

6. Hold onto [TEMP/CAL] for 5 seconds until numbers blinks.

7. Press [TEMP/CAL] to adjust readings go up to match the solution value, or Press [HOLD/MODE] to adjust readings go down to the solution value.

8. Press [POWER/ENTER] for number to set.

9. Wait until "End" stops blinking to take out of solution.

10. The meter's reading should match the solution's value now.

*** Please check or calibrate the meter monthly at room temperature..